Initial operation
Rail clamp mounting point
GA 5752 2300 GB 26
27
3
3.3.3
Mounting septic fluid jar caps with integrated overflow protection
NOTE
When mounting the aspirator to an equipment rail, attach a tube with an 8 mm
internal diameter (e.g. REF 5750 5483) using the tube adapter (REF 5752 2295)
between the hydrophobic bacterial and viral filter (1) and the opening (2) on the
back of the aspirator in order to prevent the hydrophobic filter from colliding with
the wall/equipment rail.
11
22
Fig. 11: Inserting the hydrophobic filter
Insert the conical connection (1) of the
hydrophobic filter (1) into the opening (2) on
the back of the device.
3.4
Rail clamp mounting point
Containers with an equipment mount may be attached to the rail clamp mount.
Connection to equipment rail with rail clamp (REF 5752 2048)
11
22
33
Fig. 12: Attaching rail clamp
Attaching the rail clamp
Attach rail clamp (1) to equipment rail (2).
Lock rail clamp with the locking screw (3).
3.4.1
Mounting the vacuum shift to the rail clamp
The vacuum shift is used to switch between two septic fluid jars.