Initial operation
Mounting the tubes
32
GA 5752 2300 GB 26
3
3.6.5
Tube connection for vacuum shift (REF 5752 2049)
The mounting of the tubes is described using the septic fluid jar cap (REF 5750 5362) as an
example.
3
1
2
Fig. 20: Mounting the tubes
Connect the connection tube (1) to the
lower adapter (2) of the vacuum shift.
Attach the other end of the connection tube
to the tube connector (3) on the overflow
protection device.
Overflow protection device and vacuum
shift are connected by the tube.
6
4
5
7
9
8
Fig. 21: Mounting the tubes
Attach the connection tube (4) to the left
tube connector (5) of the vacuum shift.
Attach the other end of the connection tube
to the black tube connector (6) on the
septic fluid jar cap.
Attach the connection tube (7) to the right
tube connector (8) of the vacuum shift.
Attach the other end of the connection tube
to the black tube connector (9) on the
septic fluid jar cap.
Vacuum shift is mounted.