![MEDAP 5752 4856 Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/medap/5752-4856/5752-4856_operating-instructions-manual_1759333053.webp)
Technical specifications
Weight
GA 5752 2300 GB 26
53
8
Aspirator, mobile
550 x 570 x 1080 mm
Mains cable
4000 mm
8.5
Weight
Aspirator basic equipment
ca. 11 kg
Trolley
Approx. 9.6 kg
Aspirator, mobile
ca. 20.6 kg
8.6
Vacuum depending on altitude
Altitude
Final pump vacuum Final pump vacuum
2000 m
-68 kPa
−510 mmHg
1500 m
−73 kPa
-548 mmHg
1000 m
-79 kPa
-593 mmHg
500 m
-84 kPa
-630 mmHg
0 m
-90 kPa
-675 mmHg
8.7
Electromagnetic compatibility (EMC)
The product is designed for operation in the environment specified below. The customer has to
ensure that the product is operated in an environment meeting these specifications.
8.7.1
Electromagnetic emissions
Measurement /
Standard
Compliance
Electromagnetic environment / Guidance
RF emissions
CISPR 11
Group 1
The product uses RF energy exclusively for its
own internal functions. Daher ist die HF-
Aussendung sehr gering, und es ist
unwahrscheinlich, dass benachbarte
elektronische Geräte gestört werden.
Class B
Das Produkt ist für den Gebrauch in allen
Einrichtungen einschließlich Wohnbereichen
und solchen bestimmt, die unmittelbar an ein
öffentliches Versorgungsnetz angeschlossen
sind, das auch Wohngebäude versorgt.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
Flicker emissions
IEC 61000-3-3
Complies
Tab. 8: