Operation
Suction
GA 5752 2300 GB 26
35
4
NOTE
In the event that the overflow protection device has been tripped, switch off the
aspirator. Empty and clean or replace all parts.
NOTE
Replace the hydrophobic bacterial and viral filter immediately if it is discoloured,
contaminated or oversucked. Furthermore, the filter must be exchanged if the
vacuum displayed is above −0.3 bar / 30 kPa when the vacuum controller is in the
‘max’ position and the suction tube is open.
NOTE
Rinse the suction tube briefly with clean water after each extraction cycle.
4.2.1
Switching on the aspirator
CAUTION!
Property damage!
The unit will get hot if the aspirator is switched on when vacuum is activated.
Ensure that no vacuum is present when the aspirator is switched on.
NOTE
The plug must always be accessible to ensure that the aspirator can be unplugged
from the power source at any time.
11
22
Fig. 24: Switching on the aspirator
Switch on the aspirator (1).
The green power control light (2) is
illuminated.
4.2.2
Setting the vacuum level
NOTE
Refer to the troubleshooting guide if there is insufficient vacuum or no vacuum at
all.