Initial operation
Overflow protection device
GA 5752 2300 GB 26
25
3
3.3
Overflow protection device
NOTE
If the float is not correctly seated in the overflow protection device, or no float is
used, liquid may exit in the event of oversuction.
NOTE
It is not necessary to use a hydrophobic bacterial and viral filter if a hydrophobic
bacterial and viral filter is integrated in the septic fluid jar of a disposable extraction
system.
NOTE
When using a septic fluid jar with integrated overflow protection, an additional
hydrophobic bacterial and viral filter must be used.
There are three types of overflow protection devices:
• Universal overflow protection [
Page 25]
• Surgical overflow protection [
Page 26]
• Septic fluid jar cap with integrated overflow protection [
Page 27]
3.3.1
Mounting the universal overflow protection
22
33
11
44
Fig. 8: Mounting the universal overflow protection
Engage float cage (1) with float into the lid
(2) of the overflow protection device.
Screw overflow container (3) onto the lid.
Remove cap from the connection of
equipment mount [
Page 29] and push in
equipment mount (4) of the overflow
protection device.