Maintenance
Repairs
50
GA 5752 2300 GB 26
7
22
11
Fig. 34: Replacing fuses
Remove the fuse carriers (1).
Remove the fuses (2) from the fuse
carriers.
Insert new fuses.
Insert fuse holders and screw closed.
7.4
Repairs
The following issues may require repairs by the manufacturer or an authorised service partner:
• Liquid has penetrated the device.
• The performance has significantly decreased.
• Inexplicable notifications appear.
• Abnormal noises occur.
• Functional faults cannot be rectified according to the measures in chapter Malfunctions and
troubleshooting [
page 47].
If defects are detected, the product may not be used any longer.
Make a note of the deficiencies and the REF number on the data plate and inform the responsible
ATMOS Service.
Inform the appropriate foreign representative outside Germany.
Observe the information in chapter Sending in the device [
page 51].