M133C Electric Power/Energy Calibrator
MEATEST
95
Operation manual
Example:
Voltage 230.5V in channel 1:
PHAR:VOLT1 230.5
Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:VOLT<n>:PHAS(?) <DNPD>
This command sets the phase angle between voltage output in channel <n> and internal reference
phase.
<DNPD>
Represents the phase angle in degrees.
If query is sent, calibrator returns the set value of the phase angle using standard exponential
format. Example: 156.3
is returned as 1.002.
Example:
PHAR:VOLT1:PHAS 0
PHAR:VOLT2:PHAS 120
PHAR:VOLT3:PHAS 240
Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:VOLT<n>:ENAB(?) <CPD> { ON | OFF }
This command enables (disables) the voltage output in channel <n>. Disabled outputs are OFF
also during OUTP ON.
ON - enables the output
OFF - disables the output
If query is sent, calibrator returns ON if the output is enabled or OFF if it is disabled.
Example:
PHAR:VOLT1:ENAB ON
Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:VOLT<n>:HARM<y>(?) <DNPD>
This command sets the level of <y> harmonic in output voltage in channel <n>
<DNPD>
Parameter represents the level of <y> harmonic expressed in %.
<n>
Parameter represents the output channel (1, 2 or 3).