![Meatest M133C Скачать руководство пользователя страница 60](http://html1.mh-extra.com/html/meatest/m133c/m133c_operation-manual_1758454060.webp)
MEATEST
M133C Electric Power/Energy Calibrator
Operation manual
60
8
Error messages
If an error occurs during the calibrator’s operation or control, error message is displayed on the
display. Errors can be caused by:
Incorrect control using the front panel, i.e. attempts to force a prohibited mode, e.g. setting an
out-of-range value, overloading of output terminals etc.,
Incorrect control using GPIB or RS-232 interface.
Calibrator’s fault.
Sample error message which appears when too large value is attempted to set up is shown bellow.
All error messages are displayed in the frame overlapped main screen.
The following table lists all error messages, their meaning and simple troubleshooting.
ID
Error message
Description
-430
Deadlocked.
Remote interface error. A command was received which
generates too much data to fit in the output buffer and the
output buffer is full. Command execution continues but
all data are lost.
-420
Unterminated.
Remote interface error. The calibrator was addressed to
talk but a command has not been received which sends
data to the output buffer.
-410
Interrupted.
Remote interface error. A command was received which
sends data to the output buffer, but the output buffer
contained data from a previous command. The output
buffer is cleared when power has been off, or after reset
command has been executed.
-363
Input buffer
overrun
Remote interface error.
-220
Invalid parameter
Remote interface error. An invalid character string was
received. Check to see if you have enclosed the character
string in single or double quotes and that the string
Содержание M133C
Страница 1: ...M133C 133Ci Electric Power Energy Calibrator Operation manual ...
Страница 2: ......