M133C Electric Power/Energy Calibrator
Operation manual
122
<DNPD>
Parameter represents the value number of pulses per selected unit.
If query is sent, calibrator returns set value using standard exponential format. Example: 400.0
i/kWh is returned as 4.002.
Example:
Number of pulses 1000 i/kVArh:
EACI:CONS 1000.0
EACI:UNIT VAR
Note:
Command switches the calibrator in Energy AC high current mode.
[SOUR]:EACI:TEST:TIME (?) <DNPD>
This command sets test timer in Energy AC high current mode.
<DNPD>
Parameter represents test time when pulses are counted.
If query is sent, calibrator returns the value of time using standard exponential format. Example:
20 is returned as 2.001.
Example:
EACI:TEST:TIME 10
Notes:
Command switches the calibrator in Energy AC high current mode.
[SOUR]:EACI:TEST:COUN(?) <DNPD>
This command sets test pulse counter in Energy AC high current mode.
<DNPD>
Parameter represents number of test pulses to be counted.
If query is sent, calibrator returns number of pulse counts using standard exponential format.
Example: 20 is returned as 2.001.
Example:
EACI:TEST:COUN 10
Notes:
Command switches the calibrator in Energy AC high current mode.
[SOUR]:EACI:TEST:FREQ? <DNPD>
This query returns actual frequency generated by tested energy meter.
<DNPD>
Parameter represents measured frequency.