[SOUR]:MODE?
This command returns selected mode of the calibrator { PAC | PDC | PACI | PDCI | PACE | PDCE
| PHAR | PIH | PDIP | EAC | EDC | EACI | EDCI | VAC | VDC | CAC | CDC | CACI | CDCI }.
PAC – Power AC mode
PDC – Power DC mode
PACI – Power AC high current mode
PDCI – Power DC high current mode
PACE – Power AC extended mode
PDCE – Power DC extended mode
PHAR – Power harmonic mode
PIH – Power interharmonic mode
PDIP – Power Dip/Swell mode
EAC – Energy AC mode
EDC – Energy DC mode
EACI – Energy AC high current mode
EDCI – Energy DC high current mode
VAC – Voltage AC mode
VDC – Voltage DC mode
CAC – Current AC mode
CDC – Current DC mode
CACI – Current AC high current mode
CDCI – Current DC high current mode
Example:
MODE?
[SOUR]:PAC:POW(?) <DNPD>
This command sets the power value in Power AC mode.
<DNPD>
Parameter represents the value of the power expressed in actual units (W, VA, VAr).
If query is sent, calibrator returns the set value of power using standard exponential format.
Example: 100.60 W is returned as 1.002.
Example:
AC power 1000W:
PAC:POW 1000
Note:
Command switches the calibrator in Power AC mode.
[SOUR]:PAC[:POW]:UNIT(?) <CPD> { W | VA | VAR }