M133C Electric Power/Energy Calibrator
MEATEST
127
Operation manual
Notes:
Command switches the calibrator in Energy DC high current mode.
[SOUR]:EDCI:WUP:TIME (?) <DNPD>
This command sets warm-up timer in Energy DC high current mode.
<DNPD>
Parameter represents wrm-up time when pulses are counted.
If query is sent, calibrator returns the value of time using standard exponential format. Example:
20 is returned as 2.001.
Example:
EDCI:WUP:TIME 10
Notes:
Command switches the calibrator in Energy DC high current mode.
[SOUR]:EDCI:WUP:COUN(?) <DNPD>
This command sets warm-up pulse counter in Energy DC high current mode.
<DNPD>
Parameter represents number of warm-up pulses to be counted.
If query is sent, calibrator returns number of pulse counts using standard exponential format.
Example: 20 is returned as 2.001.
Example:
EDCI:WUP:COUN 10
Notes:
Command switches the calibrator in Energy DC high current mode.
[SOUR]:VAC:VOLT(?) <DNPD>
This command sets the output voltage value in Voltage AC mode.
<DNPD>
Parameter represents the value of the output voltage expressed in V.
If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.002.
Example:
Voltage 230.5V:
VAC:VOLT 230.5
Note:
Command switches the calibrator in Voltage AC mode.