![Meatest M133C Скачать руководство пользователя страница 117](http://html1.mh-extra.com/html/meatest/m133c/m133c_operation-manual_1758454117.webp)
M133C Electric Power/Energy Calibrator
MEATEST
117
Operation manual
Note:
Command switches the calibrator in Energy DC mode.
[SOUR]:EDC:TEST:TIME (?) <DNPD>
This command sets test timer in Energy DC mode.
<DNPD>
Parameter represents test time when pulses are counted.
If query is sent, calibrator returns the value of time using standard exponential format. Example:
20 is returned as 2.001.
Example:
EDC:TEST:TIME 10
Notes:
Command switches the calibrator in Energy DC mode.
[SOUR]:EDC:TEST:COUN(?) <DNPD>
This command sets test pulse counter in Energy DC mode.
<DNPD>
Parameter represents number of test pulses to be counted.
If query is sent, calibrator returns number of pulse counts using standard exponential format.
Example: 20 is returned as 2.001.
Example:
EDC:TEST:COUN 10
Notes:
Command switches the calibrator in Energy DC mode.
[SOUR]:EDC:TEST:FREQ? <DNPD>
This query returns actual frequency generated by tested energy meter.
<DNPD>
Parameter represents measured frequency.
Calibrator returns frequency using standard exponential format. Example: 20 is returned as
2.001.
Example:
EDC:TEST:FREQ?
Notes:
Command switches the calibrator in Energy DC mode.
Содержание M133C
Страница 1: ...M133C 133Ci Electric Power Energy Calibrator Operation manual ...
Страница 2: ......