3.
Setting the current
The main power value can be changed by changing the current.
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the current
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing mA or A
softkey. The value can be set using cursor buttons or rotary knob as well.
Main power value is recalculated using new set current and existing setting of voltage
and power factor.
4.
Setting the power factor (AC power only)
If W or Var is indicated, the main power value can be changed by changing the power
factor. Change of power factor does not change the output apparent power.
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the power
factor (phase) value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing Lead or Lag
softkey. The value can be set using cursor buttons or rotary knob as well.
Main power value is recalculated using new set power factor and existing setting of
current and voltage. The calculation is only made if active or reactive power is
displayed.