MEATEST
M133C Electric Power/Energy Calibrator
Operation manual
50
7.2
Access to the calibration procedure
Calibration password is required to access the calibration procedure.
Press Menu to open the Main menu.
Select the Calibration submenu and press the „Select“ softkey.
Calibrator requests the entry of the calibration password.
Enter the correct calibration code using numeric keyboard and press ENTER.
If incorrect calibration code is entered, an error message appears on the display for
approximately 3 seconds:
Bad calibration code!
If correct calibration code is entered, calibration menu with following list appears.
Use
and
cursor buttons to move the cursor through the list of calibration items.
1.
Channel #1
Calibration data of the phase channel 1.
2.
Channel #2
Calibration data of the phase channel 2.
3.
Channel #3
Calibration data of the phase channel 3.