7
Calibration mode
The calibrator includes calibration procedure, which allows recalibration of the calibrator.
Recalibration can be performed using the buttons and menu in the calibrator.
7.1
Calibration principles
The instrument can be calibrated:
completely, i.e. all functions are calibrated in all recommended points
partially, i.e. only selected functions are calibrated in all recommended points
partially, i.e. only selected functions are calibrated in selected points
Complete calibration consists of all partial calibrations performed in the order defined by the
calibration menu. If an item of the calibration menu, e.g. „Voltage DC#1” is selected, it is not
necessary to calibrate all ranges defined by the calibration algorithm. If new calibration of all
ranges is not possible (e.g. the required standard is not available), old calibration data can be
confirmed, i.e. current step of the calibration can be skipped.
AC voltage (Voltage AC#x)
calibration is performed by setting 2 fixed points for all voltage
ranges. The frequency is 55 Hz.
AC current (Current AC#x)
calibration is performed by setting 2 fixed points for all current
ranges. The frequency is 55 Hz.
DC voltage (Voltage DC#x)
calibration is performed by setting 2 fixed points for all voltage
ranges and for both polarities (+ and -).
DC current (Current DC#x)
calibration is performed by setting 2 fixed points for all current
ranges and for both polarities (+ and -).
AC modulation voltage (Voltage MOD#x, model M133C only)
calibration is performed by
setting 2 fixed points for all voltage ranges. The frequency is 55 Hz. This calibration values are
used in P Harmonic and P Interharmonic function modes.
AC modulation current (Current MOD#x, model M133C only)
calibration is performed by
setting 2 fixed points for all current ranges. The frequency is 55 Hz. This calibration values are
used in P Harmonic and P Interharmonic function modes.
Multimeter (Meter)
calibration consist in calibration of ranges 10V DC, 20mA DC and 10 kHz.
Calibration interruption can be performed in any point of the calibration
procedure. However this particular calibration influences parameters of the
calibrator.
Accuracy of the calibrator is guaranteed when full calibration was done.