![Meatest M133C Скачать руководство пользователя страница 87](http://html1.mh-extra.com/html/meatest/m133c/m133c_operation-manual_1758454087.webp)
M133C Electric Power/Energy Calibrator
MEATEST
87
Operation manual
<DNPD>
Parameter represents the value of the power expressed in actual units (W, VA, VAr).
If query is sent, calibrator returns the set value of power using standard exponential format.
Example: 100.60 W is returned as 1.002.
Example:
AC power 1000W:
PACI:POW 1000
Note:
Command switches the calibrator in Power AC high current mode.
[SOUR]:PACI[:POW]:UNIT(?) <CPD> { W | VA | VAR }
This command selects units for power in the Power AC high current mode.
If query is sent, calibrator returns actual units in Power AC high current mode.
Example:
Select unit W for Power AC high current mode:
PACI:UNIT W
Note:
Command switches the calibrator in Power AC high current mode.
[SOUR]:PACI:VOLT(?) <DNPD>
This command sets the output voltage value.
<DNPD>
Parameter represents the value of the output voltage expressed in V.
If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.002.
Example:
Voltage 230.5V:
PAC:VOLT 230.5
Note:
Command switches the calibrator in Power AC high current mode.
[SOUR]:PACI:CURR(?) <DNPD>
This command sets the output current value in Power AC high current mode.
<DNPD>
Parameter represents the value of the output current expressed in A.
If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 A is returned as 1.001.
Содержание M133C
Страница 1: ...M133C 133Ci Electric Power Energy Calibrator Operation manual ...
Страница 2: ......