MEATEST
M133C Electric Power/Energy Calibrator
Operation manual
114
Notes:
Command switches the calibrator in Energy AC mode.
[SOUR]:EAC:WUP:TIME (?) <DNPD>
This command sets warm-up timer in Energy AC mode.
<DNPD>
Parameter represents warm-up time when pulses are counted.
If query is sent, calibrator returns the value of time using standard exponential format. Example:
20 is returned as 2.001.
Example:
EAC:WUP:TIME 10
Notes:
Command switches the calibrator in Energy AC mode.
[SOUR]:EAC:WUP:COUN(?) <DNPD>
This command sets warm-up pulse counter in Energy AC mode.
<DNPD>
Parameter represents number of warm-up pulses to be counted.
If query is sent, calibrator returns number of pulse counts using standard exponential format.
Example: 20 is returned as 2.001.
Example:
EAC:WUP:COUN 10
Notes:
Command switches the calibrator in Energy AC mode.
[SOUR]:EDC:ENER?
This query returns the set value of energy using standard exponential format. Example: 100.60 Ws
is returned as 1.002.
Note:
Query switches the calibrator in Energy DC mode.
[SOUR]:EDC:DEV?
This query returns deviation between tested energy meter end the calibrator. Calculation of
deviation is based on pulses generated by energy meter. Example: 0.612 % is returned as
6.120000e-001.