MEATEST
M133C Electric Power/Energy Calibrator
Operation manual
106
[SOUR]:PDIP:VOLT<n>:CHAN(?) <DNPD>
This command sets the level of change (Dip/Swell - t3 period) for voltage output in channel <n>.
<DNPD>
Represents the voltage in V.
If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.002.
Example:
Dip/Swell change 23.05V in channel 1:
PDIP:VOLT1:CHAN 23.05
Note:
Command switches the calibrator in Power Dip/Swell mode.
[SOUR]:PDIP:CURR<n>(?) <DNPD>
This command sets the output current value in channel <n>
<DNPD>
Parameter represents the value of the output current expressed in A.
<n>
Parameter represents the output channel (1, 2 or 3).
If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 A is returned as 1.001.
Example:
Current 23.05A in channel 1:
PDIP:CURR1 23.05
Note:
Command switches the calibrator in Power Dip/Swell mode.
[SOUR]:PDIP:CURR<n>:PHAS(?) <DNPD>
This command sets the phase angle between current output in channel <n> and internal reference
phase.
<DNPD>
Represents the phase angle in degrees.
If query is sent, calibrator returns the set value of the phase angle using standard exponential
format. Example: 156.3
is returned as 1.002.
Example:
PDIP:CURR1:PHAS 0
PDIP:CURR2:PHAS 120
PDIP:CURR3:PHAS 240