MEATEST
M133C Electric Power/Energy Calibrator
Operation manual
34
* energy counting method (control mode)
* uncertainty of energy and deviation
Set desired energy parameters using numeric keyboard, rotary knob or cursor buttons. Output
power is not yet connected to the output terminals.
Connect the instrument to be calibrated to VOLTAGE OUTPUT HI-LO and CURRENT
OUTPUT HI-LO terminals or short the CURRENT OUTPUT HI-LO terminals.
Select the Control mode.
Press ON button.
Green LED is lit above the “On” button to indicate the connection of simulated electrical
power to the output terminals. The calibrator starts energy dose generation. Main value
displays running energy, control mode parameters displays running time, counted pulses or
measured frequency. Current terminals are disconnected as soon as energy batch is finished.
Voltage terminals are disconnected only if parameter “Maintain Voltage Signal” is set to
“Off”. Otherwise the voltage terminals are disconnected after pressing “OFF” button.
Ways of energy counting (Control)
Calibration of Energy meter:
Содержание M133C