M133C Electric Power/Energy Calibrator
MEATEST
89
Operation manual
If query is sent, calibrator returns the value of frequency using standard exponential format.
Example: 50Hz is returned as 5.001.
Example:
Frequency 60Hz:
PACI:FREQ 60
Notes:
Command switches the calibrator in Power AC high current mode.
[SOUR]:PDCI:POW(?) <DNPD>
This command sets the power value in Power DC high current mode.
<DNPD>
Parameter represents the value of the power expressed in W.
If query is sent, calibrator returns the set value of power using standard exponential format.
Example: 100.60 W is returned as 1.002.
Example:
DC power 1000W:
PDCI:POW 1000
Note:
Command switches the calibrator in Power DC high current mode.
[SOUR]:PDCI:VOLT(?) <DNPD>
This command sets the output voltage value in Power DC high current mode.
<DNPD>
Parameter represents the value of the output voltage expressed in V.
If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.002.
Example:
Voltage 230.5V:
PDCI:VOLT 230.5
Note:
Command switches the calibrator in Power DC high current mode.
[SOUR]:PDCI:CURR(?) <DNPD>
This command sets the output current value in Power DC high current mode.
<DNPD>
Parameter represents the value of the output current expressed in A.