M133C Electric Power/Energy Calibrator
MEATEST
115
Operation manual
Note:
Query switches the calibrator in Energy DC mode.
[SOUR]:EDC:POW?
This query returns the set value of power using standard exponential format. Example: 100.60 W
is returned as 1.002.
Note:
Query switches the calibrator in Energy DC mode.
[SOUR]:EDC:VOLT(?) <DNPD>
This command sets the output voltage value.
<DNPD>
Parameter represents the value of the output voltage expressed in V.
If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.002.
Example:
Voltage 230.5V:
EDC:VOLT 230.5
Note:
Command switches the calibrator in Energy DC mode.
[SOUR]:EDC:CURR(?) <DNPD>
This command sets the output current value in Energy DC mode.
<DNPD>
Parameter represents the value of the output current expressed in A.
If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 A is returned as 1.001.
Example:
Current 23.05A:
EDC:CURR 23.05
Note:
Command switches the calibrator in Energy DC mode.
[SOUR]:EDC:CONT(?) <CPD> { PACK | CNT1 | CNT2 | TIM1 | TIM2 | FR1 | FR2 |
FR3}
This command selects energy counting control mode.