80
20.12 Calibration - Flow Calibration
Midas Simulator State Chart
Function
Selection -
Zero Gas
Calibration
After A/D Readings
are completed
Accept
Read Flow
Zero Offset
Prepare For
Calibration
Accept
Cancel
Cancel
Calibration - Flow Calibration
Enter
Setpoint 2
Flow
Adjust
Setpoint 1
Flow
Adjust
Setpoint 2
Flow
Enter
Setpoint 1
Flow
Read A/D For
Setpoint 1
Read A/D For
Setpoint 2
Accept (Check)
Up
Down
Cancel (X)
Calibration
Abort changes
When adjusting flow, increases flow (simulated
on flow slider). When entering flow, increments
value (1 second hold for faster increment).
When adjusting flow, decreases flow (simulated
on flow slider). When entering flow, decrements
value (1 second hold for faster decrement).
Additional Button Functions
Accept
Accept
Accept
Accept
After A/D Readings
are completed
After A/D Readings
are completed
Cancel
Cancel
Cancel
Cancel
Cancel
Cancel
Notes:
1) Setpoint 1 values:
Target Flow: 350 cc/min
Minimum Flow: 300 cc/min
Maximum Flow: 400 cc/min
2) Setpoint 2 values:
Target Flow: 650 cc/min
Minimum Flow: 600 cc/min
Maximum Flow: 700 cc/min
3) For “Adjust Flow” states, flow would be adjust to approximately
the target flow rate using the arrow buttons and would be
measured on an external flowmeter.
4) For “Enter Flow” states, the actual flow shown on the external
flowmeter would be entered.
5) For “Read Flow” states, the display will count down the read
process. Final values would be saved after the Read Setpoint 2
Flow state.
Содержание midas
Страница 1: ...1 TOXIC FLAMMABLE AND OXYGEN GAS DETECTOR OP E RAT ING INSTRUCTIONS ...
Страница 44: ...44 6 Remove the 4 pump module screws 7 Remove the two fixing clips and disconnect the tubes at the manifold ...
Страница 55: ...55 Diagram 40 IP Address setting in Windows XP Diagram 41 Sample MIDAS web page ...
Страница 57: ...57 15 2 Modbus TCP Installation 15 3 Power over Ethernet POE Installation ...