67
Finishing
Wait for the gas reading to return to normal levels. This may take some time, especially if concentrations above
the recommended levels have been used or if the test gas has been applied for an extended period of time.
Ensure that any latched faults or alarms that have been generated by the testing have been cleared.
Make sure that the gas detection system is fully restored to its normal operating state and that any control system
overrides used during the testing have been removed.
The alarms generated by the testing will have created entries in the transmitter’s history log.
Troubleshooting
If the test does not produce a satisfactory result, check the following points.
1. Check the gas cylinder concentration and type are correct
2. Check the expiration date of the cylinder.
3. Check that there is sufficient gas left in the cylinder.
4. Check that there are no leaks on the test system.
5. Make sure that the tubing length is as short as possible and that there are no blockages.
6. Make sure that the transmitter is not displaying a fault before the test is started.
19 Warranty Statement
All products are designed and manufactured to the latest internationally recognized standards by Zellweger
Analytics under a Quality Management System that is certified to ISO 9001.
As such, this instrument (including the pump) is warranted under proper use, to the original end-user purchaser,
against defects in materials and workmanship within 24 months from the date of first turn-on. Separate warranty
conditions apply to the sensor cartridges limited as indicated below. During this period, Zellweger Analytics will
repair or replace defective parts on an exchange basis, F.O.B. to approved service centers on a global basis.
This warranty does not cover damage caused by accident, abuse, abnormal operating conditions or extreme
poisoning of the sensor cartridge.
Defective goods must be returned by the Buyer at the Buyer’s own expense to Zellweger Analytics premises
accompanied by a detailed report stating the nature of the defect. Returned goods must detail the Service Event
Number (SE#) clearly on the package and the Buyer shall obtain an SE# by contacting Zellweger Analytics in
advance.
If no such report is included then Zellweger Analytics reserves the right to charge an investigative fee (prices
available upon request) before any repair or replacement is performed.
Zellweger Analytics shall not be liable for any loss or damage whatsoever or howsoever occasioned which may
be a direct or indirect result of the use or operation of the Contract Goods by the Buyer or any Party.
This warranty covers the gas detector and parts sold to the Buyer only by authorized distributors, dealers
and representatives as appointed by Zellweger Analytics. A warranty claim will only be accepted if a proof of
purchase is submitted and all conditions obtained within this Warranty are met. When in the opinion of Zellweger
Analytics, a warranty claim is valid, Zellweger Analytics will repair or replace the defective product free of charge
and send it or any replacement back.
The initial warranty period is not extended by virtue of any works carried out there under.
The decision to repair or replace parts shall be determined by Zellweger Analytics.
Содержание midas
Страница 1: ...1 TOXIC FLAMMABLE AND OXYGEN GAS DETECTOR OP E RAT ING INSTRUCTIONS ...
Страница 44: ...44 6 Remove the 4 pump module screws 7 Remove the two fixing clips and disconnect the tubes at the manifold ...
Страница 55: ...55 Diagram 40 IP Address setting in Windows XP Diagram 41 Sample MIDAS web page ...
Страница 57: ...57 15 2 Modbus TCP Installation 15 3 Power over Ethernet POE Installation ...