9.2 PUMP REPLACEMENT
The pump module has been designed to allow easy replacement. New pump modules are supplied with new
springs, bracket and tubing pre-assembled for quick release / replacement
Diagram 26. Location of pump module
The following procedure should be followed carefully and only performed by suitably trained personnel.
1. Isolate the power to the detector.
2. Unscrew the thumbscrew located on the front panel.
3. Remove the cover by pulling it forward off the main chassis.
4. Unscrew the two retaining screws located at the bottom front of the chassis.
5. Pull the main chassis forward to disconnect it from the mounting bracket assembly.
Pump module