Chapter 6
Advanced protein features
Page 6-6
MAN 0487
Running a large number of sub-runs sequentially significantly increases the risk of
Joule heating of the sample, so the protein mobility measurement is split into the
smaller
groups
of sub-runs with a delay between groups to allow the sample to
relax.
Size measurement
Full details of the size measurement window will be found in the main Zetasizer
Nano user manual. A brief overview follows.
Settings
Description
Perform a size
measurement...
With the
Perform a size measurement before and after
the Protein mobility measurement
check box selected a
size measurement will also be performed as well as the
standard zeta potential measurement.
Angle of detection
The measurement angle is fixed to 13 degree forward scat-
ter. This is to ensure that the size is measured at the same
angle as the zeta potential.
Measurement
duration
The
Measurement
duration
setting can affect the accuracy
and repeatability of the size results.
Measurements
This option allows a sample to be measured more than once;
to investigate the effect on particle size over time, or to prove
repeatability.