Chapter 5
Microrheology
Page 5-10
MAN 0487
Sample preparation advice
The correct preparation of the tracer and sample elements for the measurement is
important to ensure that reliable and repeatable results are achieved.
The concentration of tracer samples used will depend to a great extent on the
concentration of your sample in the dispersant.
Following are some important considerations for the preparation.
Zeta potential measurements
For the initial tests of the zeta potential the following starting concentrations are
suggested.
Tracer in dispersant/buffer (or solvent) only
0.5
μ
l tracer particles to 10ml sample (0.05
μ
l/ml).
Tracer in dispersant/buffer (or solvent) including sample
Add 10-100
μ
l of the tracer in dispersant to 1ml of the sample in dispersant.
Start at the low end and increase the concentration if required.
Size measurements - to hide sample scattering
Size measurements are performed to check that the tracer particle's scattering
completely hides the sample scattering; therefore the size of the tracer particle
should be larger than the sample.