Chapter 3
Surface zeta potential cell
Page 3-8
MAN 0487
The cuvette must not be filled more than the recommended maximum depth
of 20mm before insertion of the cell
.
Tilt the cuvette to a maximum angle of 45°
. This is to avoid spilling the dis-
persant.
Slowly insert the cell into the cuvette until the sample holder, barrel and elec-
trodes are covered
. As the cell is inserted it displaces the sample so any bub-
bles will be pushed out from the top of the electrode gap.
Once the electrodes are covered bring the cuvette up to the vertical
.
Inspect the combined cell and cuvette and check for any bubbles
. If bubbles
are present around the electrode or nylon screw, gently tap the bottom of the
cuvette to dislodge these. If not dislodged repeat the above sequence.
ill 8799
Stop
Stop
20mm
max
max
45°
1
2
3
4
6
5
B
A
A