General cells and cuvettes
Chapter 2
Zetasizer Nano accessories guide
Page 2-23
ill 8509
Place the Dip cell immediately into an empty cuvette
for storage.
This will prevent any potential damage occuring either to the cell electrodes or
the workspace.
Remove the sample cuvette afterwards and place in the cuvette holder
.
Note
Storage is also provided in the Dip cell case if the cell is not to be used for a
while.
Cleaning the Dip cell
Clean the cell thoroughly between measurements, especially between different
types of sample. Cross-contamination between samples can seriously affect the
results.
Caution!
Do
not
immerse the complete cell. Only the sample electrodes must dip in
to the dispersant, as shown below.
Clean the Dip cell electrodes regularly. They are made of solid palladium and can
be cleaned physically and chemically. Follow the instructions below:
Immerse the electrodes in a gentle ultrasound bath (30 Watts) for five to 15
minutes before use. Use the dispersant used for the previous sample as the
cleaning fluid. If this dispersant contains additives such as surfactants, follow
this by ultrasonicating for two minutes in the pure solvent.
1
2
3
Tap