Chapter 2
General cells and cuvettes
Page 2-24
MAN 0487
ill 6763
Warning!
Take care: ultrasonication can produce a fine aerosol of the bath liquid.
Remove the electrodes from the bath and rinse them with pure solvent. A pipe
cleaner can be used for
gentle
cleaning of electrodes.
To protect the Dip cell after cleaning, we recommend placing it in an empty
cuvette for storage.
Before making a measurement, rinse the electrodes and cuvette with the sample to
be measured.
When changing the sample, thoroughly rinse the electrodes with pure dispersant.
Note
The electrode holder is made from Natural PEEK (Polyetheretherketone)
which is resistant to a wide range of chemical products. However, seek
advice from Malvern and the sample manufacturer before using strong
acid or base.
Chemical compatibility - Dip cell
With proper use, only the central electrode section of the Dip cell will ever come in
contact with sample. The outer components of the Dip cell will only come into
contact if spillage or overfilling occurs.
Component
Materials
Central section
Electrode casing
Electrodes
Natural PEEK (Polyetheretherketone)
Palladium
Outer components
Top and side casing
Contacts
Natural PEEK (Polyetheretherketone)
Phosphor Bronze with Nickel plating