General cells and cuvettes
Chapter 2
Zetasizer Nano accessories guide
Page 2-27
External surfaces of the assembled cell can be wiped clean with a weak soap solu-
tion.
Intensive cleaning
The cell first has to be disassembled before cleaning can be performed.
ill 8447
Remove the screw cap
.
Separate the two halves of the cell
by pulling the rear casing vertically away
from the metal front.
Note how the electrode chambers and quartz measurement cell block are
assembled
.
Remove the chambers and cell block from the metal front casing
.
Detach the pipework and remove the top port
.
Protect the cell block from damage
.
Once the cell has been disassembled, cleaning can be performed as described in the
following table.
1
2
3
4
5
6