Chapter 5
Microrheology
Page 5-16
MAN 0487
With the measurement is complete, an intensity plot should also be displayed
showing one peak. The software compares the two measured size results and tests
for the following:
The size of the tracer particle should be
larger
than the sample particles
The tracer particle's scattering should completely hide that of the sample
scattering. If the sample scattering is still visible the software will prompt for
more tracer to be added to the sample.
The tracer scattering must obscure the sample scattering by more than 90%, but
not so much that the sample becomes turbid, therefore making any subsequent
measurement difficult. When the software is satisfied with the comparison between
the two size measurements, the microrheology measurement can be started.
A dialogue will appear indicating the next step (the microrheology
measurement) can be performed. Press next to start the microrheology
measurement automatically.
Microrheology measurement
This measurement is a single size measurement from which the MSD and
rheological parameters are calculated.
The microrheology measurement display
The microrheology measurement displays are virtually identical to those shown
when performing a standard zeta potential or size measurement.
These standard tabs are explained in the main Zetasizer Nano user manual.
Содержание Zetasizer Nano Series
Страница 2: ......
Страница 3: ...Zetasizer Nano accessories guide MAN0487 Issue 1 1 April 2013...
Страница 12: ...Chapter 1 Introduction and accessory range Page 1 6 MAN 0487...
Страница 42: ...Chapter 2 General cells and cuvettes Page 2 30 MAN 0487...
Страница 98: ...Chapter 5 Microrheology Page 5 20 MAN 0487...
Страница 125: ......