Chapter 2
General cells and cuvettes
Page 2-16
MAN 0487
Note
The stoppers
must
be fitted before a measurement is performed. Ensure
that one stopper is fitted firmly, and the other one loosely, to avoid pressur-
isation of the cell.
Inserting the folded capillary cell
ill 8507
Place a
thermal
contact
plate
into the recess on either side of the folded cap-
illary cell
. The plates provide increased temperature stability.
Open the cell area lid by pushing the button
in front of the lid.
1
2
3
4
Plates MUST be
positioned correctly
Raised side of
plate against
sunken side
of cell
Flat side of
plate against
flat side of cell