The Magstim Company Ltd.
©
67
NOP08-EN-10
5.2 Non-Recoverable Error
The non-recoverable screen (right) is
displayed when the system detects an error/
failure condition that impacts the essential
performance of the system.
If this screen is displayed during the use of
electrocautery, the error condition may be
recoverable by switching the nerve monitor
off using the mains power switch on the
rear of the base, waiting 30 seconds and
switching the nerve monitor back on.
External
USB
If the error persists, this is a non-recoverable error
and you must contact the service provider to
arrange for a repair.
The alarm banner displays the type of fault
detected and the service contact details are
displayed on the screen. In order to export the fault
details insert a USB memory stick into the external
USB port (right) and press
. This information
can then be sent to the relevant service provider
for further evaluation.
5.3 System Alarms & Messages
Banner Message
Action
'Check Electrode Connection'
Check the electrode's connection to the patient. Re-
insert or replace the electrode if necessary.
Check that the electrode lead is correctly connected
in the pre-amplifier socket.
This alarm may be shown briefly during the use of
electrocautery, if the problem persists contact your
service provider. Contact details can be found on page
ii.
' Impedance Too Low. Reduce
Primary Output Voltage'
Ensure that the stimulating probe is not in contact
with a metallic object.
Reduce the stimulating voltage by rotating the left
stimulator (STIM1) dial clockwise.
If the problem persists contact your service provider.
Contact details can be found on page ii.
Note:
DO NOT switch the nerve monitor off & on to clear an error message, unless
instructed to do so in the following table:
Содержание NEUROSIGN V4
Страница 3: ...The Magstim Company Ltd iii NOP08 EN 10 Welcome to the Neurosign V4 Intraoperative Nerve Monitor...
Страница 4: ...The Magstim Company Ltd iv NOP08 EN 10 Page intentionally left blank...
Страница 6: ...The Magstim Company Ltd vi NOP08 EN 10 Page intentionally left blank...
Страница 8: ...The Magstim Company Ltd viii NOP08 EN 10 Page intentionally left blank...
Страница 10: ...The Magstim Company Ltd 10 NOP08 EN 10 Page intentionally left blank...
Страница 12: ...The Magstim Company Ltd 12 NOP08 EN 10 Page intentionally left blank...
Страница 13: ...The Magstim Company Ltd 13 NOP08 EN 10 SECTION 1 PRODUCT OVERVIEW...
Страница 18: ...The Magstim Company Ltd 18 NOP08 EN 10 Page intentionally left blank...
Страница 19: ...The Magstim Company Ltd 19 NOP08 EN 10 SECTION 2 ACCESSORIES...
Страница 25: ...The Magstim Company Ltd 25 NOP08 EN 10 SECTION 3 WARNINGS SYMBOLS...
Страница 35: ...The Magstim Company Ltd 35 NOP08 EN 10 SECTION 4 INSTRUCTIONS FOR USE...
Страница 65: ...The Magstim Company Ltd 65 NOP08 EN 10 SECTION 5 TROUBLE SHOOTING...
Страница 74: ...The Magstim Company Ltd 74 NOP08 EN 10 Page intentionally left blank...
Страница 75: ...The Magstim Company Ltd 75 NOP08 EN 10 SECTION 6 CARE MAINTENANCE...
Страница 79: ...The Magstim Company Ltd 79 NOP08 EN 10 SECTION 7 TECHNICAL INFORMATION...
Страница 86: ...www neurosign com The Magstim Company Ltd...