4.9 Pre-Amplifier Replacement
To change the pre-amplifier mid-surgery do not press stop to end monitoring. If the
pre-amplifier is changed mid-surgery to a higher capacity (i.e. 4 to 8) transfer the
electrodes to the corresponding sockets, the channels do not have to be reconfigured.
If the pre-amplifier is changed mid-surgery to a lower capacity (i.e. 8 to 4) the screen
below will be shown:
Figure 4.10. Monitoring - Pre-Amplifier Replaced
• In this case it is not necessary to match the channel colours. Instead connect the
required electrodes into any available channel on the pre-amplifier. The reference
electrode/s (green) must be transferred to the reference electrode channel.
• Connect the pre-amplifier and press 'YES' to reconfigure the channels from the
channel settings tab (see "c. Procedure Settings" on page 49 to reconfigure your
channels). Do not stop monitoring.
• Press 'NO' to end the procedure and exit the monitoring screen.
Содержание NEUROSIGN V4
Страница 3: ...The Magstim Company Ltd iii NOP08 EN 10 Welcome to the Neurosign V4 Intraoperative Nerve Monitor...
Страница 4: ...The Magstim Company Ltd iv NOP08 EN 10 Page intentionally left blank...
Страница 6: ...The Magstim Company Ltd vi NOP08 EN 10 Page intentionally left blank...
Страница 8: ...The Magstim Company Ltd viii NOP08 EN 10 Page intentionally left blank...
Страница 10: ...The Magstim Company Ltd 10 NOP08 EN 10 Page intentionally left blank...
Страница 12: ...The Magstim Company Ltd 12 NOP08 EN 10 Page intentionally left blank...
Страница 13: ...The Magstim Company Ltd 13 NOP08 EN 10 SECTION 1 PRODUCT OVERVIEW...
Страница 18: ...The Magstim Company Ltd 18 NOP08 EN 10 Page intentionally left blank...
Страница 19: ...The Magstim Company Ltd 19 NOP08 EN 10 SECTION 2 ACCESSORIES...
Страница 25: ...The Magstim Company Ltd 25 NOP08 EN 10 SECTION 3 WARNINGS SYMBOLS...
Страница 35: ...The Magstim Company Ltd 35 NOP08 EN 10 SECTION 4 INSTRUCTIONS FOR USE...
Страница 65: ...The Magstim Company Ltd 65 NOP08 EN 10 SECTION 5 TROUBLE SHOOTING...
Страница 74: ...The Magstim Company Ltd 74 NOP08 EN 10 Page intentionally left blank...
Страница 75: ...The Magstim Company Ltd 75 NOP08 EN 10 SECTION 6 CARE MAINTENANCE...
Страница 79: ...The Magstim Company Ltd 79 NOP08 EN 10 SECTION 7 TECHNICAL INFORMATION...
Страница 86: ...www neurosign com The Magstim Company Ltd...