6.4 Servicing & Device Lifetime
a. Electrical Testing
The nerve monitor and associated accessories must be tested annually for electrical
safety in accordance with local regulation or hospital policy. There is an equipotential
point fitted near the power switch which may be used for electrical safety testing, or
connected to a theatre equipotential earthing system.
Note:
Once powered down, you must wait a minimum of 30 seconds before starting
the system up again to allow the nerve monitor to restart correctly.
b. Servicing
All servicing and repair work must be carried out by The Magstim Company Ltd. or an
authorised service personnel designated by the company, Service provider contact
details can be found on page ii.
Excepting of those stated in "6.3 User Maintenance & Calibration" on page 76,
no user-serviceable parts are found within the Neurosign
®
V4 Intraoperative Nerve
Monitor. Opening the case presents a serious threat of electric shock and immediate
risk to the user. Any attempt to remove the security screws will invalidate the product
guarantee.
Performance verification should be carried out by service personnel annually, once
the guarantee period has expired.
A service manual is available on request for those who have undergone service training
on the equipment and to suitably authorised personnel.
b. Damage Inspection
Before use the equipment and accessories should be checked for any damage or
tampering as described below:
• Check the cables for splits/cracks
• Check the pre-amplifier and stimulator pod for cracks in the case and for any
signs of tampering
If the equipment is found to be damaged or faulty discontinue use immediately and
contact your local Neurosign
®
representative to arrange for a repair or replacement.
c. Storage Management
See "e. Storage Management" on page 60 for information on storage management,
d. Battery Voltage
The battery voltage can be found on the system details screen.
To access the system details screen;
• From the 'home' screen, press
and then press
• The battery voltage can be found under 'Display Unit' (top left).
Содержание NEUROSIGN V4
Страница 3: ...The Magstim Company Ltd iii NOP08 EN 10 Welcome to the Neurosign V4 Intraoperative Nerve Monitor...
Страница 4: ...The Magstim Company Ltd iv NOP08 EN 10 Page intentionally left blank...
Страница 6: ...The Magstim Company Ltd vi NOP08 EN 10 Page intentionally left blank...
Страница 8: ...The Magstim Company Ltd viii NOP08 EN 10 Page intentionally left blank...
Страница 10: ...The Magstim Company Ltd 10 NOP08 EN 10 Page intentionally left blank...
Страница 12: ...The Magstim Company Ltd 12 NOP08 EN 10 Page intentionally left blank...
Страница 13: ...The Magstim Company Ltd 13 NOP08 EN 10 SECTION 1 PRODUCT OVERVIEW...
Страница 18: ...The Magstim Company Ltd 18 NOP08 EN 10 Page intentionally left blank...
Страница 19: ...The Magstim Company Ltd 19 NOP08 EN 10 SECTION 2 ACCESSORIES...
Страница 25: ...The Magstim Company Ltd 25 NOP08 EN 10 SECTION 3 WARNINGS SYMBOLS...
Страница 35: ...The Magstim Company Ltd 35 NOP08 EN 10 SECTION 4 INSTRUCTIONS FOR USE...
Страница 65: ...The Magstim Company Ltd 65 NOP08 EN 10 SECTION 5 TROUBLE SHOOTING...
Страница 74: ...The Magstim Company Ltd 74 NOP08 EN 10 Page intentionally left blank...
Страница 75: ...The Magstim Company Ltd 75 NOP08 EN 10 SECTION 6 CARE MAINTENANCE...
Страница 79: ...The Magstim Company Ltd 79 NOP08 EN 10 SECTION 7 TECHNICAL INFORMATION...
Страница 86: ...www neurosign com The Magstim Company Ltd...