5.4 Troubleshooting (Physical)
Note:
DO NOT switch the nerve monitor off & on to clear an error, unless instructed to
do so in the following table:
Problem
Cause
Action
Pre-amplifier audio
connection test fails
during the use of
electrocautery
The use of electrocautery
during the pre-amplifier
audio connection test
Refer to "i. Test failure" on page 40.
Signs of external
electrical interference
External electrical
interference
Connect the Earthing Cable to the
equipotential point. See "5.5 Neurosign V4
Blank screen
Lack of power
Check the power lead is correctly installed
in the power input socket on the rear of
the base and that the power switch and
mains plug switch are on.
Display connection cable is
not properly connected.
Check the cable is properly connected and
secured at both ends (see "1.4 Rear View"
If the problem persists please contact your
service provider (page ii).
Touch screen not
responding correctly.
The touch screen needs
recalibrating.
Refer to "6.3 User Maintenance &
If the problem persists please contact your
service provider (page ii).
Touch screen not
responding at all.
DCU failure
Contact your service provider (page ii).
The stimulating probe
fails to stimulate and
the current confirm
does not light.
Stimulator fail
Check probe is properly connected to the
primary stimulator connection (STIM 1).
Replace probe.
If the problem persists please contact your
service provider (page ii).
Содержание NEUROSIGN V4
Страница 3: ...The Magstim Company Ltd iii NOP08 EN 10 Welcome to the Neurosign V4 Intraoperative Nerve Monitor...
Страница 4: ...The Magstim Company Ltd iv NOP08 EN 10 Page intentionally left blank...
Страница 6: ...The Magstim Company Ltd vi NOP08 EN 10 Page intentionally left blank...
Страница 8: ...The Magstim Company Ltd viii NOP08 EN 10 Page intentionally left blank...
Страница 10: ...The Magstim Company Ltd 10 NOP08 EN 10 Page intentionally left blank...
Страница 12: ...The Magstim Company Ltd 12 NOP08 EN 10 Page intentionally left blank...
Страница 13: ...The Magstim Company Ltd 13 NOP08 EN 10 SECTION 1 PRODUCT OVERVIEW...
Страница 18: ...The Magstim Company Ltd 18 NOP08 EN 10 Page intentionally left blank...
Страница 19: ...The Magstim Company Ltd 19 NOP08 EN 10 SECTION 2 ACCESSORIES...
Страница 25: ...The Magstim Company Ltd 25 NOP08 EN 10 SECTION 3 WARNINGS SYMBOLS...
Страница 35: ...The Magstim Company Ltd 35 NOP08 EN 10 SECTION 4 INSTRUCTIONS FOR USE...
Страница 65: ...The Magstim Company Ltd 65 NOP08 EN 10 SECTION 5 TROUBLE SHOOTING...
Страница 74: ...The Magstim Company Ltd 74 NOP08 EN 10 Page intentionally left blank...
Страница 75: ...The Magstim Company Ltd 75 NOP08 EN 10 SECTION 6 CARE MAINTENANCE...
Страница 79: ...The Magstim Company Ltd 79 NOP08 EN 10 SECTION 7 TECHNICAL INFORMATION...
Страница 86: ...www neurosign com The Magstim Company Ltd...