The Magstim Company Ltd.
©
69
NOP08-EN-10
Banner Message
Action
' System Storage Critical'
Contact your service provider to relieve the system
memory log.
Contact details can be found on page ii.
' Connection to Pre-Amplifier
Lost'
Disconnect the pre-amplifier lead from the pre-
amplifier connection socket on the rear of the screen.
Reconnect the pre-amplifier, if the error remains
contact your service provider. Contact details can be
found on page ii.
' Connection to Stimulator Pod
Lost'
Disconnect the stimulator-pod lead from the pre-
amplifier.
Reconnect the stimulator-pod, if the error remains
contact your service provider. Contact details can be
found on page ii.
' Battery Voltage Low'
Contact your service provider to replace the internal
battery. Contact details can be found on page ii.
EMG Audio is Muted
Un-mute the audio by rotating the master volume dial
clockwise, or briefly pressing the dial in.
' Pre-Amplifier Failure.
Disconnect and Contact
Service'
Disconnect the pre-amplifier lead from the pre-
amplifier connection socket on the rear of the screen
and contact your service provider. Contact details can
be found on page ii.
' Report Storage Critical'
The user report storage is full.
See "e. Storage Management" on page 60 for
details on how to relieve the user storage.
Содержание NEUROSIGN V4
Страница 3: ...The Magstim Company Ltd iii NOP08 EN 10 Welcome to the Neurosign V4 Intraoperative Nerve Monitor...
Страница 4: ...The Magstim Company Ltd iv NOP08 EN 10 Page intentionally left blank...
Страница 6: ...The Magstim Company Ltd vi NOP08 EN 10 Page intentionally left blank...
Страница 8: ...The Magstim Company Ltd viii NOP08 EN 10 Page intentionally left blank...
Страница 10: ...The Magstim Company Ltd 10 NOP08 EN 10 Page intentionally left blank...
Страница 12: ...The Magstim Company Ltd 12 NOP08 EN 10 Page intentionally left blank...
Страница 13: ...The Magstim Company Ltd 13 NOP08 EN 10 SECTION 1 PRODUCT OVERVIEW...
Страница 18: ...The Magstim Company Ltd 18 NOP08 EN 10 Page intentionally left blank...
Страница 19: ...The Magstim Company Ltd 19 NOP08 EN 10 SECTION 2 ACCESSORIES...
Страница 25: ...The Magstim Company Ltd 25 NOP08 EN 10 SECTION 3 WARNINGS SYMBOLS...
Страница 35: ...The Magstim Company Ltd 35 NOP08 EN 10 SECTION 4 INSTRUCTIONS FOR USE...
Страница 65: ...The Magstim Company Ltd 65 NOP08 EN 10 SECTION 5 TROUBLE SHOOTING...
Страница 74: ...The Magstim Company Ltd 74 NOP08 EN 10 Page intentionally left blank...
Страница 75: ...The Magstim Company Ltd 75 NOP08 EN 10 SECTION 6 CARE MAINTENANCE...
Страница 79: ...The Magstim Company Ltd 79 NOP08 EN 10 SECTION 7 TECHNICAL INFORMATION...
Страница 86: ...www neurosign com The Magstim Company Ltd...