The Magstim Company Ltd.
©
38
NOP08-EN-10
Note:
Ensure the AC mains lead is connected to the
power input socket (
19
, page 39 before use but do not
switch the unit on yet.
First, clean and/or protect all equipment according to
hospital procedure (see page 76 for information on
cleaning and sterilising).
a. Connection
Once the patient is anaesthetised, the electrodes may
be inserted into the relevant locations on the patient.
Plug the electrodes into the corresponding sockets on
the pre-amplifier (
22
, or
22 & 24
, right).
Plug the green reference electrode/s into either green
socket (
23
, right).
Plug the pre-amplifier lead (
25
, right) into the correct
socket on the rear of the screen (
12
, page 39).
Attach the pre-amplifier to the side rail of the operating
table using the mounting clip.
Plug one end of the stimulator pod lead into the socket
on the stimulator pod (
29
, right) and connect the other
end to the socket on the pre-amplifier (
25
, right).
25. Stimulator
Pod Socket
22. Electrode
Channels
1 - 4
24. Electrode
Channels
5 - 8
23. Reference
Electrode
Channel
21. Mounting
Clip
26. Pre-Amplifier
Connection Lead
Primary
Stimulator
Sockets
Stimulator
Pod Socket
Mounting
Clip
27.
28.
29.
4.2 Connection & Disconnection
Attach the stimulator pod to the side rail
of the operating table (away from the
surgical zone) using the mounting clip.
Set the power switch to ON (I) (
20
, page
39), The monitor will run a connection
and audio test (see "b. Connection &
Note:
It is advised not to drape the patient
until after the electrode impedance
check.
If used, attach the mute sensor to the
electrosurgical unit’s monopolar pencil
lead and plug the mute sensor into the
mute sensor socket
(
15
, Page 39).
Note:
It is advised that you do not connect
the stimulating probe/s until required.
8-Channel variant shown
Содержание NEUROSIGN V4
Страница 3: ...The Magstim Company Ltd iii NOP08 EN 10 Welcome to the Neurosign V4 Intraoperative Nerve Monitor...
Страница 4: ...The Magstim Company Ltd iv NOP08 EN 10 Page intentionally left blank...
Страница 6: ...The Magstim Company Ltd vi NOP08 EN 10 Page intentionally left blank...
Страница 8: ...The Magstim Company Ltd viii NOP08 EN 10 Page intentionally left blank...
Страница 10: ...The Magstim Company Ltd 10 NOP08 EN 10 Page intentionally left blank...
Страница 12: ...The Magstim Company Ltd 12 NOP08 EN 10 Page intentionally left blank...
Страница 13: ...The Magstim Company Ltd 13 NOP08 EN 10 SECTION 1 PRODUCT OVERVIEW...
Страница 18: ...The Magstim Company Ltd 18 NOP08 EN 10 Page intentionally left blank...
Страница 19: ...The Magstim Company Ltd 19 NOP08 EN 10 SECTION 2 ACCESSORIES...
Страница 25: ...The Magstim Company Ltd 25 NOP08 EN 10 SECTION 3 WARNINGS SYMBOLS...
Страница 35: ...The Magstim Company Ltd 35 NOP08 EN 10 SECTION 4 INSTRUCTIONS FOR USE...
Страница 65: ...The Magstim Company Ltd 65 NOP08 EN 10 SECTION 5 TROUBLE SHOOTING...
Страница 74: ...The Magstim Company Ltd 74 NOP08 EN 10 Page intentionally left blank...
Страница 75: ...The Magstim Company Ltd 75 NOP08 EN 10 SECTION 6 CARE MAINTENANCE...
Страница 79: ...The Magstim Company Ltd 79 NOP08 EN 10 SECTION 7 TECHNICAL INFORMATION...
Страница 86: ...www neurosign com The Magstim Company Ltd...