Problem
Cause
Action
The stimulating probe
fails to stimulate, but
the current confirm
does light.
Stimulator fail
Check current level is adequate.
Check stimulator pod connection.
If the problem persists please contact your
service provider (page ii).
Pre-Amplifier
Replaced
The pre-amplifier has been
replaced mid-surgery for one
of a lower capacity.
See "4.8 Reconfiguring Channels" on page
54.
Silence/ Lack of sound.
Sound is muted
Increase the main volume by rotating the
master volume dial clockwise.
Check the individual channel volume is
not muted or too low by looking for the
channel volume in the top right of each
channel window (see page 46). Press
the channel volume and use the slider to
increase.
Silence when stimulating;
the device has not detected
any voltage in the muscle,
the muscle has not
contracted.
The muscle may not contract
for a variety of reasons;
the nerve may not be
stimulated, the nerve may
be damaged or transected,
or any neuromuscular block
may be too deep.
The surgeon in charge should decide
whether or not to proceed with the
procedure taking the possible causes into
consideration.
Monitor failure.
If you feel the silence is an indication of
a monitoring failure please contact your
service provider (page ii).
Waveforms are too
large and cannot be
clearly seen.
Amplitude scale is set too
low.
Increase amplitude scale (see "i. Channel
Waveforms appear
and immediately
disappear
Waveform threshold is set
too low.
Increase the waveform threshold (see "i.
Содержание NEUROSIGN V4
Страница 3: ...The Magstim Company Ltd iii NOP08 EN 10 Welcome to the Neurosign V4 Intraoperative Nerve Monitor...
Страница 4: ...The Magstim Company Ltd iv NOP08 EN 10 Page intentionally left blank...
Страница 6: ...The Magstim Company Ltd vi NOP08 EN 10 Page intentionally left blank...
Страница 8: ...The Magstim Company Ltd viii NOP08 EN 10 Page intentionally left blank...
Страница 10: ...The Magstim Company Ltd 10 NOP08 EN 10 Page intentionally left blank...
Страница 12: ...The Magstim Company Ltd 12 NOP08 EN 10 Page intentionally left blank...
Страница 13: ...The Magstim Company Ltd 13 NOP08 EN 10 SECTION 1 PRODUCT OVERVIEW...
Страница 18: ...The Magstim Company Ltd 18 NOP08 EN 10 Page intentionally left blank...
Страница 19: ...The Magstim Company Ltd 19 NOP08 EN 10 SECTION 2 ACCESSORIES...
Страница 25: ...The Magstim Company Ltd 25 NOP08 EN 10 SECTION 3 WARNINGS SYMBOLS...
Страница 35: ...The Magstim Company Ltd 35 NOP08 EN 10 SECTION 4 INSTRUCTIONS FOR USE...
Страница 65: ...The Magstim Company Ltd 65 NOP08 EN 10 SECTION 5 TROUBLE SHOOTING...
Страница 74: ...The Magstim Company Ltd 74 NOP08 EN 10 Page intentionally left blank...
Страница 75: ...The Magstim Company Ltd 75 NOP08 EN 10 SECTION 6 CARE MAINTENANCE...
Страница 79: ...The Magstim Company Ltd 79 NOP08 EN 10 SECTION 7 TECHNICAL INFORMATION...
Страница 86: ...www neurosign com The Magstim Company Ltd...