4.8 Reconfiguring Channels
a. Reconfiguring Channels Mid-Surgery
In some cases it may necessary to reconfigure the procedure channels mid-surgery.
The process for this is described below.
To re-configure the channels mid-surgery:
• From the monitoring menu, press
• Select the Channel Setting tab.
• Refer to "c. Procedure Settings" on page 49 to reconfigure you channels.
b. Configuring an 8-channel Procedure For a 4-channel Pre-Amplifier
This functionality may be used when you wish to configure an 8-channel procedure
before surgery, which is not supported by the connected 4-channel pre-amplifier as
standard.
Note: In cases where the monitoring is active, follow the instructions in section 4.9.
• Select or create a surgeon profile (see "a. Profile Management"). The procedure
management screen will then be displayed.
• Toggle 'Include Unsupported Procedures' ON ( ) at the top of the screen.
• Unsupported procedures can now be seen with the
symbol.
• Select the desired procedure and press
to duplicate that procedure. Type in
a name for this new procedure setup and press OK.
• This new procedure setup can now be edited by pressing
.
• To configure the new procedure for use with the connected pre-amplifier, refer to
"c. Procedure Settings" on page 49 to reconfigure your channels.
Содержание NEUROSIGN V4
Страница 3: ...The Magstim Company Ltd iii NOP08 EN 10 Welcome to the Neurosign V4 Intraoperative Nerve Monitor...
Страница 4: ...The Magstim Company Ltd iv NOP08 EN 10 Page intentionally left blank...
Страница 6: ...The Magstim Company Ltd vi NOP08 EN 10 Page intentionally left blank...
Страница 8: ...The Magstim Company Ltd viii NOP08 EN 10 Page intentionally left blank...
Страница 10: ...The Magstim Company Ltd 10 NOP08 EN 10 Page intentionally left blank...
Страница 12: ...The Magstim Company Ltd 12 NOP08 EN 10 Page intentionally left blank...
Страница 13: ...The Magstim Company Ltd 13 NOP08 EN 10 SECTION 1 PRODUCT OVERVIEW...
Страница 18: ...The Magstim Company Ltd 18 NOP08 EN 10 Page intentionally left blank...
Страница 19: ...The Magstim Company Ltd 19 NOP08 EN 10 SECTION 2 ACCESSORIES...
Страница 25: ...The Magstim Company Ltd 25 NOP08 EN 10 SECTION 3 WARNINGS SYMBOLS...
Страница 35: ...The Magstim Company Ltd 35 NOP08 EN 10 SECTION 4 INSTRUCTIONS FOR USE...
Страница 65: ...The Magstim Company Ltd 65 NOP08 EN 10 SECTION 5 TROUBLE SHOOTING...
Страница 74: ...The Magstim Company Ltd 74 NOP08 EN 10 Page intentionally left blank...
Страница 75: ...The Magstim Company Ltd 75 NOP08 EN 10 SECTION 6 CARE MAINTENANCE...
Страница 79: ...The Magstim Company Ltd 79 NOP08 EN 10 SECTION 7 TECHNICAL INFORMATION...
Страница 86: ...www neurosign com The Magstim Company Ltd...