7. Enter the sensors actual value (in this case depth) in the “
Actual value
” field.
8. Click on the “
Accept
” button to calculate the new slope and offset parameter values for the associated
channel. Clicking on the “
Cancel
” button terminates the procedure without altering the parameter
values.
For a short video of the procedure click "
Play
"-
MACE XCi User Manual - 215 -
Содержание HVFlo XCi
Страница 1: ......
Страница 10: ...MACE XCi User Manual 10...
Страница 18: ...MACE XCi User Manual 18...
Страница 21: ...MACE XCi User Manual 21...
Страница 50: ...MACE XCi User Manual 50...
Страница 98: ...MACE XCi User Manual 98...
Страница 173: ...MACE XCi User Manual 173...
Страница 178: ...MACE XCi User Manual 178...
Страница 191: ...NOTE In the example the Net flowrate will be the sum of Flow Rate 1 Flow Rate 2 Flow Rate 3 MACE XCi User Manual 191...
Страница 211: ...For a short video of the procedure click Play MACE XCi User Manual 211...
Страница 233: ...5 Tighten all three screws 6 Reconnect the battery cable assembly to the backplane board MACE XCi User Manual 233...
Страница 238: ...3 Unscrew the locking ring using the tool and remove completely MACE XCi User Manual 238...
Страница 239: ...4 Remove the damaged ceramic depth sensor MACE XCi User Manual 239...
Страница 250: ...MACE XCi User Manual 250...