"2-Point Calibration" function
To correctly calibrate an environmental sensor it is important to calibrate it over its useful range. For example,
if a depth sensor has a range of 0 - 2 metres but you only wish to use it to measure the range 0.0 metres to
1.25 metres, you should re-calibrate it across the reduced range.
1. Click “
2-point calibration
” from the “
Edit channel
” dialogue box and the “
2 point calibration
”
dialogue box appears.
The “
Current value
” from the sensor (in this case a depth sensor with frequency output) is displayed
at the top of the window and is updated approximately once per second.
2. To carry out a two point calibration, prepare the measurement set up for the first calibration point
(either the sensors upper or lower range limit).
3. When the “
Current value
” has stabilised, click on the “
Set
” button in the “
1st point
” box. The
“
Current value
” is copied to the “
Measured value
” field of the “
1st point
” box and the “
Actual
value
” field is enabled.
4. Enter the sensors actual value (in this case depth) in the “
Actual value
” field.
5. Prepare the measurement set up for the second calibration point (the opposite sensor range limit from
that used in point 2 above). For example, if you used the upper range limit in point 2, now use the
lower range limit.
6. When the “
Current value
” has stabilised, click on the “
Set
” button in the “
2nd point
” box. The
“
Current value
” is copied to the “
Measured value
” field of the “
2nd point
” box and the “
Actual
value
” field is enabled.
MACE XCi User Manual - 214 -
Содержание HVFlo XCi
Страница 1: ......
Страница 10: ...MACE XCi User Manual 10...
Страница 18: ...MACE XCi User Manual 18...
Страница 21: ...MACE XCi User Manual 21...
Страница 50: ...MACE XCi User Manual 50...
Страница 98: ...MACE XCi User Manual 98...
Страница 173: ...MACE XCi User Manual 173...
Страница 178: ...MACE XCi User Manual 178...
Страница 191: ...NOTE In the example the Net flowrate will be the sum of Flow Rate 1 Flow Rate 2 Flow Rate 3 MACE XCi User Manual 191...
Страница 211: ...For a short video of the procedure click Play MACE XCi User Manual 211...
Страница 233: ...5 Tighten all three screws 6 Reconnect the battery cable assembly to the backplane board MACE XCi User Manual 233...
Страница 238: ...3 Unscrew the locking ring using the tool and remove completely MACE XCi User Manual 238...
Страница 239: ...4 Remove the damaged ceramic depth sensor MACE XCi User Manual 239...
Страница 250: ...MACE XCi User Manual 250...