Edit a "Depth" - channel type
1. Enter the appropriate "
Slope
" and "
Offset
" values that are either supplied with the depth sensor or
have been calculated.
2. Perform either an "
Offset adjust
" or "
2-point calibration
" to ensure accurate measuring results are
recorded
NOTE: See either
or
"2-Point Calibration" function
for
full details of these procedures.
3. Click "
Apply
" to save the settings and
"Exit
" to return to the "
Configure channels
" dialogue box.