Symptom
Cause
Remedy
"Sensor
error"
on
LCD
in
log
file
in "Measure & Display"
Sensor (SM) Bus error
Doppler sensor not plugged into
card
Check connections
Corrosion in sensor connector
Check for corrosion and replace
if necessary
Sensor cable damaged
Check for physical damage,
cuts in cable etc.
Faulty sensor
Follow
and if
damaged
Velocity readings are neg-
ative
Sensor is installed facing down-
stream
of
sensor
Sensor is installed facing down-
stream
Change
in
Velocity range not set correctly
Ensure
set to
MACE XCi User Manual - 225 -