![Mace HVFlo XCi Скачать руководство пользователя страница 164](http://html1.mh-extra.com/html/mace/hvflo-xci/hvflo-xci_product-manual_3203897164.webp)
"Doppler stream index"- channel type
The Doppler stream index provides a simple measure of the number of acoustically reflective particles and
their relative signal strengths. This index is generated for each velocity calculation. As a general rule of thumb
an index value of 10 or greater is a requisite for repeatable velocity readings.
WARNING: If an index value of less than 10 is consistently observed the site may
not be suitable for Doppler ultrasonic velocity measurement
NOTE: There are no editable parameters for this channel type.
MACE XCi User Manual - 164 -