Fill Height:
The maximum depth measured by the EchoFlo. Refer to the EchoFlo Quick Start Guide for
further details.
Sensor Span:
Is the difference between the sensor output at the top of the “
Fill Height
” (2000Hz) and the
sensor output at the bottom (976Hz).The frequency output of an EchoFlo has a span of 1024Hz .
Manually calculating Offset
Offset = - (Slope x Sensor output at zero)
Metric example -
you wish to calculate the slope and offset for an EchoFlo that has been configured with a
“
Fill Height
” of 0.75m (75cm).
The following calculations will provide the sensor’s slope and offset:
Slope
= 0.75 ÷ (2000-976)
= 0.75 ÷ 1024
= 0.00073
Offset
= - (0.00073 x 976)
= - 0.71484
US example -
you wish to calculate the slope and offset for an EchoFlo that has been configured with a “
Fill
Height
” of 30”.
The following calculations will provide the sensor’s slope and offset:
Slope
= 30 ÷ (2000-976)
= 30 ÷ 1024
= 0.02929
Offset
= - (0.02929 x 976)
= - 28.59375
NOTE: Be sure to record the relevant values for “
Slope
”,“
Offset
”, and depth units.
MACE XCi User Manual - 167 -
Содержание HVFlo XCi
Страница 1: ......
Страница 10: ...MACE XCi User Manual 10...
Страница 18: ...MACE XCi User Manual 18...
Страница 21: ...MACE XCi User Manual 21...
Страница 50: ...MACE XCi User Manual 50...
Страница 98: ...MACE XCi User Manual 98...
Страница 173: ...MACE XCi User Manual 173...
Страница 178: ...MACE XCi User Manual 178...
Страница 191: ...NOTE In the example the Net flowrate will be the sum of Flow Rate 1 Flow Rate 2 Flow Rate 3 MACE XCi User Manual 191...
Страница 211: ...For a short video of the procedure click Play MACE XCi User Manual 211...
Страница 233: ...5 Tighten all three screws 6 Reconnect the battery cable assembly to the backplane board MACE XCi User Manual 233...
Страница 238: ...3 Unscrew the locking ring using the tool and remove completely MACE XCi User Manual 238...
Страница 239: ...4 Remove the damaged ceramic depth sensor MACE XCi User Manual 239...
Страница 250: ...MACE XCi User Manual 250...