Symptom
Cause
Remedy
Integration period not long
enough to collect data
Increase "
Insufficient acoustic targets at
site
Read "
"
guidelines
My meter IS reading flow
when it's NOT flowing!
Normally occurs due to moving
water - leaky gate, foot valve
Check seals of gates and valves
Entrapped aquatic life in pipe
Adjust "
" using
Water "sloshing" in pipe around
sensor level
Make sure water level recedes
fully, or pipe stays completely
full
Electromagnetic interference
(EMI) causing false "flow"
Find source of EMI and rectify if
possible
MACE XCi User Manual - 221 -
Содержание HVFlo XCi
Страница 1: ......
Страница 10: ...MACE XCi User Manual 10...
Страница 18: ...MACE XCi User Manual 18...
Страница 21: ...MACE XCi User Manual 21...
Страница 50: ...MACE XCi User Manual 50...
Страница 98: ...MACE XCi User Manual 98...
Страница 173: ...MACE XCi User Manual 173...
Страница 178: ...MACE XCi User Manual 178...
Страница 191: ...NOTE In the example the Net flowrate will be the sum of Flow Rate 1 Flow Rate 2 Flow Rate 3 MACE XCi User Manual 191...
Страница 211: ...For a short video of the procedure click Play MACE XCi User Manual 211...
Страница 233: ...5 Tighten all three screws 6 Reconnect the battery cable assembly to the backplane board MACE XCi User Manual 233...
Страница 238: ...3 Unscrew the locking ring using the tool and remove completely MACE XCi User Manual 238...
Страница 239: ...4 Remove the damaged ceramic depth sensor MACE XCi User Manual 239...
Страница 250: ...MACE XCi User Manual 250...