Installation with a mounting plate
DO NOT cut the sensor cable. Re-termination of the sensor is highly specialised and MUST be carried out at
a MACE technical facility.
A mounting plate is normally used in large pipes where a cost-effective solution is required. The sensor is
attached to the plate, which is then bolted to the inside surface of the pipe. This plate is supplied in
polypropylene.
1. Attach the sensor head securely to the mounting plate by the six 3mm screws supplied (M3 x 8 mm
Phillips head countersunk 304 s/s).
WARNING: Use only a hand screwdriver. Do not over tighten these screws as
permanent damage to the thread may result
2. Tie the cable to the mounting plate using the holes and cable ties provided. Make sure that the knob
of the cable tie sits to the back of the assembly, out of the flow.
DO NOT OVERTIGHTEN CABLE
TIES.
3. The cable from the sensor to the XCi device should then be run inside the pipe and secured to the pipe
in such a manner as to prevent damage by debris in the pipe. The cable should be routed through
electrical conduit.
MACE XCi User Manual - 48 -
Содержание HVFlo XCi
Страница 1: ......
Страница 10: ...MACE XCi User Manual 10...
Страница 18: ...MACE XCi User Manual 18...
Страница 21: ...MACE XCi User Manual 21...
Страница 50: ...MACE XCi User Manual 50...
Страница 98: ...MACE XCi User Manual 98...
Страница 173: ...MACE XCi User Manual 173...
Страница 178: ...MACE XCi User Manual 178...
Страница 191: ...NOTE In the example the Net flowrate will be the sum of Flow Rate 1 Flow Rate 2 Flow Rate 3 MACE XCi User Manual 191...
Страница 211: ...For a short video of the procedure click Play MACE XCi User Manual 211...
Страница 233: ...5 Tighten all three screws 6 Reconnect the battery cable assembly to the backplane board MACE XCi User Manual 233...
Страница 238: ...3 Unscrew the locking ring using the tool and remove completely MACE XCi User Manual 238...
Страница 239: ...4 Remove the damaged ceramic depth sensor MACE XCi User Manual 239...
Страница 250: ...MACE XCi User Manual 250...