"Offset adjust" function
This function is used to “fine tune” a sensor reading whenever a sensor is deployed or reading inaccurately.
For example, a depth sensor may be sitting slightly off the bottom of the stream, or the user has deployed it
above a silt line on the side of a pipe.
1. Click “
Offset adjust
” from the “
Edit channel
” dialogue box and the “
Channel offset adjust
”
dialogue box appears.
The “
Current value
” from the sensor (in this case a depth sensor with frequency output) is displayed
at the top of the window and is updated approximately once per second.
2. To adjust the “
Current value
”, enter your current measured value in the “
Corrected value
” field.
3. Click on the “
Accept
” button to calculate the new “
Offset
” parameter value for the associated channel.
Clicking on the “
Cancel
” button terminates the procedure without altering the offset value.
For a short video of the procedure click "
Play
"-
-
WARNING: When calibrating a "
Depth
" channel, measurements should ALWAYS
be carried out from the bottom of the channel/pipe to the top of the water level
MACE XCi User Manual - 213 -
Содержание HVFlo XCi
Страница 1: ......
Страница 10: ...MACE XCi User Manual 10...
Страница 18: ...MACE XCi User Manual 18...
Страница 21: ...MACE XCi User Manual 21...
Страница 50: ...MACE XCi User Manual 50...
Страница 98: ...MACE XCi User Manual 98...
Страница 173: ...MACE XCi User Manual 173...
Страница 178: ...MACE XCi User Manual 178...
Страница 191: ...NOTE In the example the Net flowrate will be the sum of Flow Rate 1 Flow Rate 2 Flow Rate 3 MACE XCi User Manual 191...
Страница 211: ...For a short video of the procedure click Play MACE XCi User Manual 211...
Страница 233: ...5 Tighten all three screws 6 Reconnect the battery cable assembly to the backplane board MACE XCi User Manual 233...
Страница 238: ...3 Unscrew the locking ring using the tool and remove completely MACE XCi User Manual 238...
Страница 239: ...4 Remove the damaged ceramic depth sensor MACE XCi User Manual 239...
Страница 250: ...MACE XCi User Manual 250...