5.5 Error messages
(in alphabetical order)
Carattere non valido
You entered an invalid character or comma (such as , \ / : * ? “ < > |). All the
fields only allow integer values (except
numbers separated by dot or comma -decimal mark-).
Disconnected cable
This message is allowed to appear only when the current output is selected
(
= CURRENT). +Iout cable is not connected or connected
incorrectly or broken. Check the connection. The error is further signalled
through the FAULT output. For more information refer to the “4.9 Fault output”
section on page 20.
Encoder HW counts to be converted into the selected output range must
be lower than the maximum total counts
While programming the encoder using the PROG mode (see on pages 47 and
50), the operator has confirmed in the parameter
Encoder HW counts
in
Programming mode: Prog (single ramp)
a number of information for the travel greater than the
maximum number of HW information available for the encoder. The maximum
number of information (
) which can be set is 67 108 864 (2
26
).
If you confirm a value greater than 2
26
this alarm is invoked to appear.
Error creating the file
If an error occurs while saving the encoder configuration to file (see the “5.4.1
Saving the encoder configuration to a file” section on page 54), then this error
message is displayed.
Error reading the file
If an error occurs while uploading an encoder configuration file (see the “5.4.2
Loading the encoder configuration from a file” section on page 55), then this
error message is displayed.
Value not valid: preset greater than the maximum position
The operator has entered next to the
PRESET
box a
value which is beyond the limits of the selected analogue range (see the